RAC Exam - EU MDDA / MDD MDR NEWEST ACTUAL EXAM
COMPLETE 130 QUESTIONS AND CORRECT DETAILED ANSWERS/
GRADE A+ ASSURED
1. Similar but more detailed than the Essential Requirements: General Safety and
Performance Requirements
2. This ISO standard includes the most recent update to specific medical device
companies quality management system?: 13485:2016
3. Which entity affixes it's identification number near the CE Mark?: Notified Body, if
involved in the conformity assessment
4. Un-classified devices not requiring CE mark: Custom-made, Investigational,
Compassionate/Orphan
5. All ----- contributing to the intended purpose must be considered: Modes of action
6. Define transient use: less than 60 minutes
7. Define short-term use: between 60 minutes and 30 days
8. A material is considered to have a biological effect if it actively and intentionally
----: induces, alters, or prevents a response from tissues that is mediated by specific
reactions at a molecular level
9. Refers to a material's degradation within the body and metabolic elimination of the
resulting degradation products from the body: Absorption
10. Combination devices are most often classified as: Class III
11. MEDDEV 2.4/1^7 is a guidance for: Classification under MDD
12. Standalone software is considered what type of device: Active
13. Software that drives a device or influences its use automatically falls in to which
classification?: That of the device
14. If a manufacturer and an NB cannot agree on the classification who is consulted
for a solution?: Competent Authority
15. How can a manufacturer appeal the CA's decision of a classification?: in the national
courts
16. Define horizontal/Level I standards: General Standards relevant to all/wide range of
product types
17. Define Semi-horizontal/level 2 standards: requirements applicable to a range of
similar products (surgical instruments)
18. Define vertical/level 3 standards: apply to a single product type or narrow range
of products
19. Define comparator device: established device bearing a CE Mark that is used as a
reference in a clinical investigation NOT a predicate device like FDA)
20. What is the ISO for biocompatibility?: ISO 10993
21. IEC 60601 regards -: electrical safety and electric magnetic disturbance
22. The amount of a medical device's testing is/should be commensurate with
, -: it's potential RISK to human health
23. Name the two forms of clinical data used in CE marking: 1. Compilation of relevant
scientific literature currently available on the device's intended purpose, techniques
employed and critical evaluation of the compilation
2. Results and conclusions of a specifically designed clinical investigation
24. What is contained in Annex 1 of the MDD, AIMDD and MDR?: Essential
Requirements (gsprs in MDR)
25. A clinical evaluation is required for -: every device, Class I through III under all
directives
26. If the literature shows new or other risks what may need to be updated?: -
Risk management plan
27. What is not required if a manufacturer can collect clinical evidence from
literature?: Clinical investigation
28. State of the art =: current generally accepted technology and performance output
not to be confused with top-of-the-line.
29. FSCA: Field Safety Corrective Action
30. ISO 14155: Clinical investigations of medical devices in human subjects
31. IB: Investigators brochure - compilation of the current clinical and non-clinical
information on the investigational medical device(s) relevant to the clinical investigation.
32. CIR: Clinical Investigation report
33. Whish entity is legally responsible for compliance with regulations and the first
point of contact for EU authorities?: Authorised Representative (as applicable, if not the
manufacturer itself)
34. Which entity is liable for defective products in the post market phase: Authorised rep
35. EMA: European Medicines Agency
36. Define EC type-examination: NB body reviews and, at times, performs device testing
37. The post market surveillance concept is concept is addressed in these two ISO
standards: 13485 and 14971
38. The criteria and procedures used by mfr's, CA's and other interested parties to
notify and handle incidents and FSCA's or recalls are known col- lectively as the -:
Medical Device Vigilance System
39. GPSD: General Product Safety Directive
40. What are the entities that make up the "economic operators"?: manufactur- ers,
Authorised Reps, importers, and distributors
41. When is a PMCF study mandatory? (MDD): When CE marking is based solely on clinical
data from equivalent devices
42. PSUR: Periodic Safety Update Report
The benefits of buying summaries with Stuvia:
Guaranteed quality through customer reviews
Stuvia customers have reviewed more than 700,000 summaries. This how you know that you are buying the best documents.
Quick and easy check-out
You can quickly pay through credit card or Stuvia-credit for the summaries. There is no membership needed.
Focus on what matters
Your fellow students write the study notes themselves, which is why the documents are always reliable and up-to-date. This ensures you quickly get to the core!
Frequently asked questions
What do I get when I buy this document?
You get a PDF, available immediately after your purchase. The purchased document is accessible anytime, anywhere and indefinitely through your profile.
Satisfaction guarantee: how does it work?
Our satisfaction guarantee ensures that you always find a study document that suits you well. You fill out a form, and our customer service team takes care of the rest.
Who am I buying these notes from?
Stuvia is a marketplace, so you are not buying this document from us, but from seller samuelwachira. Stuvia facilitates payment to the seller.
Will I be stuck with a subscription?
No, you only buy these notes for $8.49. You're not tied to anything after your purchase.