Which of the following can be potential safety concerns to a patient in the static
magnetic field?
A. Electrically conductive materials
B. Ferrous objects
C. Magnetic objects
D. All of the choices. - Answer D. All of the choices
An object that retains its magnetic field when exiting the static magnetic field is
call what?
A. Paramagnetic
B. Diamagnetic
C. Ferromagnetic
D. Superparamagnetic - Answer C. Ferromagnetic
What direction is the vector of a permanent MR unit?
A. Horizontal
B. Vertical
C. Diagonal
D. Variable - Answer B. Vertical
What separates a superconducting MR unit from a resistive MR unit?
A. Cryogen
,B. Coils of wires
C. Permanent magnets
D. none of the choices - Answer A. Cryogen\
The increasing strength of magnetism as we move closer to the MR unit is called
what?
A. Ferromagnetism
B. Spatial magnetic gradient field
C. Vector
D. B0 - Answer B. Spatial magnetic gradient field
What is responsible for attracting ferrous objects into the MR unit?
A. Spatial magnetic gradient fields
B. Strong uniform magnetic fields
C. Gradient coils
D. Radiofrequency system - Answer A. Spatial magnetic gradient fields
If we placed a small ferrous object at isocenter, what would we expect to
happen?
A. The object would experience no attraction.
B. The object would immediately be attracted to the bore.
C. The object would levitate.
D. The object would be repulsed by the unit. - Answer A. The object would
experience no attraction.
Rotational force will be the strongest at what location?
A. Isocenter
B. Outside the MR room
C. At the gantry opening
D. At the 5 gauss line - Answer A. Isocenter
Passive shielding uses which of the following?
,A. Coils of wires with electrical current running through them.
B. Magnetically conductive materials
C. Plastic
D. Wood - Answer B. Magnetically conductive materials
How can we reduce risks associated with Lenz's force?
A. No risks are associated with Lenz's force.
B. Use a cold compress to reduce heating.
C. Move as quickly as possible to get the patient out of the scanner quickly.
D. Advance patient into the scanner slowly. - Answer D. Advance Patient into the
scanner slowly.
An implant has a deflection of 23 degrees. In terms of static magnetic fields, is it
safe for this patient to have an MRI?
A. Yes
B. Must call the manufacture to determine conditions.
C. No
D. Not enough information. - Answer A. Yes
Which of the following stainless steels are non-ferrous?
A. Ferritic
B. Martensitic
C. Austenitic
D. All are ferrous - Answer C. Austenitic
It is not safe for a patient to cross over the 5 gauss line. How many millitesla is
this?
A. 0.5mT
B. 5mT
C. 50mT
D. 0.005mT - Answer A. 0.5mT
, A patient is placed in a magnetic field that changes over a distance from the MR
unit. What do we need in order to meet the manufacture limits if a patient with an
implant will be safe from the threats associated with the static magnetic field
prior to their MRI.
A. Spatial gradient map.
B. Compression
C. The MRSO sole approval.
D. All of the choices. - Answer A. Spatial gradient map
The spatial gradient of 1.5 T MR unit, not the maximum spatial gradient has a
lower spatial gradient than a 3.0 T MR unit.
A. True
B. False - Answer B. False
Which of the following are documented biological effects associated with the
static magnetic field?
A. Magnetophosphenes
B. Magnetohydrodynamic Effect
C. Headaches
D. Difficulty breathing - Answer A. Magnetophosphenes, B.
Magnetohydrodynamic Effect and C. Headaches.
What is the purpose of the time-varying gradient magnetic field?
A. To produce resonance in our hydrogen nuclei
B. To change the precessional speeds of hydrogen
C. To produce net magnetism
D. To form an image. - Answer B. To change the precessional speeds of
hydrogen.
The effect of the time-varying gradient is greatest where?
A. Periphery of the patient.
B. At isocenter
The benefits of buying summaries with Stuvia:
Guaranteed quality through customer reviews
Stuvia customers have reviewed more than 700,000 summaries. This how you know that you are buying the best documents.
Quick and easy check-out
You can quickly pay through credit card or Stuvia-credit for the summaries. There is no membership needed.
Focus on what matters
Your fellow students write the study notes themselves, which is why the documents are always reliable and up-to-date. This ensures you quickly get to the core!
Frequently asked questions
What do I get when I buy this document?
You get a PDF, available immediately after your purchase. The purchased document is accessible anytime, anywhere and indefinitely through your profile.
Satisfaction guarantee: how does it work?
Our satisfaction guarantee ensures that you always find a study document that suits you well. You fill out a form, and our customer service team takes care of the rest.
Who am I buying these notes from?
Stuvia is a marketplace, so you are not buying this document from us, but from seller luzlinkuz. Stuvia facilitates payment to the seller.
Will I be stuck with a subscription?
No, you only buy these notes for $13.49. You're not tied to anything after your purchase.