If your biometry instrument has one pseudophakic eye-type setting, what is the only IOL
material that can accurately be measured? - CORRECT ANSWER- Polymethyl
methacrylate (PMMA)
If a gain setting is increased in order to obtain a ...
ROUB Study Guide UPDATED Exam
Questions and CORRECT Answers
If your biometry instrument has one pseudophakic eye-type setting, what is the only IOL
material that can accurately be measured? - CORRECT ANSWER-✔✔ Polymethyl
methacrylate (PMMA)
If a gain setting is increased in order to obtain a 100% tall retina and sclera echo, what type of
cataract is being evaluated? - CORRECT ANSWER-✔✔ A dense cataract
What transducer frequency is generally used for a biometry A-scan probe? - CORRECT
ANSWER-✔✔ 10 MHz
What is the velocity of a biometer that only uses an average velocity to measure the entire
axial length? - CORRECT ANSWER-✔✔ 1550 m/sec
What is one of the most important features of a biometer? - CORRECT ANSWER-✔✔
Display screen
What are the velocity settings for a phakic eye? - CORRECT ANSWER-✔✔ 1532 m/sec for
the aqueous; 1641 m/sec for the lens; and, 1532 m/sec for the vitreous cavity
What should the examiner do to reduce the presence of a reverberation echo created by a
dense cataract? - CORRECT ANSWER-✔✔ Reduce the gain
What is the term for the electronic calipers that measure the distance between two or more
anatomic interfaces? - CORRECT ANSWER-✔✔ Gates
Of the following settings on the biometer (gain, gates, A-scan probe frequency, eye type, and
screen brightness), which cannot be changed by the examiner? - CORRECT ANSWER-✔✔
A-scan probe frequency
What is the proper sound velocity used to measure the distance between the anterior and
posterior lens interfaces in a phakic eye? - CORRECT ANSWER-✔✔ 1641 m/sec
,When performing immersion biometry, which gate in relation to the initial spike represents
the cornea? - CORRECT ANSWER-✔✔ Corneal gate is to the right of the initial spike
What is the possible cause for a biometry instrument to indicate an unusually short
measurement of a pseudophakic eye with a PMMA implant? - CORRECT ANSWER-✔✔
Multiple signal of the echo with improper identification of the retina
What biometer setting changes the degree of echo amplification? - CORRECT ANSWER-
✔✔ Gain
What is the term for the relative strength or amplitude of an echo displayed on a biometer? -
CORRECT ANSWER-✔✔ Reflectivity
When measuring an eye that has no crystalline lens, the instrument should be set on which
eye type? - CORRECT ANSWER-✔✔ Aphakic
What is the term for the "markers" or "cursors" displayed on the screen of a biometry
instrument that provide electronic measurements of distance between two or more acoustic
interfaces? - CORRECT ANSWER-✔✔ Gates
What is the term for the capacity of an ultrasound instrument to display two or more
interfaces that are very anatomically close to one another as separate echoes? - CORRECT
ANSWER-✔✔ Resolution
How does increasing the gain setting on the biometer affect the display echo? - CORRECT
ANSWER-✔✔ It increases the amount of amplification of the echo
What is the range of ophthalmic biometry frequencies? - CORRECT ANSWER-✔✔ 8-12
MHz
How is the resolution affected when the gain is set at maximum? - CORRECT ANSWER-
✔✔ The resolution is less
, A non-focused sound beam is always utilized by which type of scan? - CORRECT
ANSWER-✔✔ Diagnostic A-scan
What scan is used to obtain a two-dimensional image of the globe? - CORRECT ANSWER-
✔✔ Diagnostic B-scan
What is the term for a one-dimensional amplitude scan? - CORRECT ANSWER-✔✔ A-scan
The strength of an echo is dependent on what three factors? - CORRECT ANSWER-✔✔
Gain, difference in density, and proper alignment
What type of scan uses Prager or scleral shells? - CORRECT ANSWER-✔✔ Immersion A-
scan
How often should the filter of a biometry instrument be vacuumed? - CORRECT ANSWER-
✔✔ Once a month
Measurement of axial length of the eye, thickness of an extraocular muscle, and tumor height
are all within the definition of what term? - CORRECT ANSWER-✔✔ Biometry
What is the leading cause of an unplanned, post-operative refractive error? - CORRECT
ANSWER-✔✔ An error in the measurement of axial length
An error of 0.1 mm in the axial length measurement can result in a post-operative refraction
error of how many diopters? - CORRECT ANSWER-✔✔ 0.25D
If multiple axial length readings are taken from a single eye, each measurement should be
within how many millimeters? - CORRECT ANSWER-✔✔ 0.2 mm
The difference in axial length between two eyes should measure within how many
millimeters? - CORRECT ANSWER-✔✔ 0.3 mm
The benefits of buying summaries with Stuvia:
Guaranteed quality through customer reviews
Stuvia customers have reviewed more than 700,000 summaries. This how you know that you are buying the best documents.
Quick and easy check-out
You can quickly pay through credit card or Stuvia-credit for the summaries. There is no membership needed.
Focus on what matters
Your fellow students write the study notes themselves, which is why the documents are always reliable and up-to-date. This ensures you quickly get to the core!
Frequently asked questions
What do I get when I buy this document?
You get a PDF, available immediately after your purchase. The purchased document is accessible anytime, anywhere and indefinitely through your profile.
Satisfaction guarantee: how does it work?
Our satisfaction guarantee ensures that you always find a study document that suits you well. You fill out a form, and our customer service team takes care of the rest.
Who am I buying these notes from?
Stuvia is a marketplace, so you are not buying this document from us, but from seller MGRADES. Stuvia facilitates payment to the seller.
Will I be stuck with a subscription?
No, you only buy these notes for $10.49. You're not tied to anything after your purchase.